ICU System Ventilator
Model: MCV-3010
Main Features:
1. Microprocessor controlled, Pneumatically driven
electronically controlled
2. 12.1" TFT touch screen, user friendly interface
3. Built-in medical air compressor, built-in backup
battery up to 3hrs
4. LCD display ventilation parameters, preset
parameters alarm indication and breathing waveform
5. Applied for adult and child
6. CE marked
Air compressor technique info
7. Output pressure: 0.4MPa ±0.1MPa
8. Noise ≤50dB
9. Power: AC 220V ± 10%, 50Hz ± 2%
Main Specification
Ventilation Mode
IPPV, SIGH, A/C (SIPPV), SIMV, MANUAL, PEEP, CPAP/SPONT
(standard)
PCV, PSV, BI-LEVEL, S,T, S/T (optional)
Parameters
●Tidal volume: 20~2000ML
●Ventilation volume per minute: 0L~20L
●Frequency: 1~100/min
●I/E Ratio: 4:1~1:8
●Holding time: 0~6s
●O2 concentration: 21~100% adjustable.
● Pressure limitation: 0.5~8.0KPa
● Safety Pressure ≤12.5KPa
● PEEP: 0~3.0KPa
● CPAP: 0~2.0KPa
● Inspiration trigger sensitivity:
pressure -2.0KPa (base on PEEP)
flow: 1~30L/min
● Sign: 1-5 (per 100 times)
Ventilation monitoring
●Tidal volume,Minute ventilation,Frequency, Airway Pressure, inspiration time, PEEP, FiO2, PCV,PSV
●Waveform of time-airway Pressure.
Alarm (Audible and visible reminding)
●No tidal volume, ventilation volume is too high (low) ,
Airway Pressure is too high (low), frequency is too high,
AC Power off , back up battery voltage is too low
Technique info
● Gas source: Oxygen, Air
● Working Pressure: 0.4Mpa
● Reserve Condition: Temperature:-10~40℃, Relative humidity: ≤90%
● Working condition: Temperature:5~40 ℃, Relative humidity:≤80%